Analizador de Oxígeno (%) - OPM 8000
Paramagnetic analyser | Analysis of O2 in % nivel
General
The OPM 8000 Analizador de Oxígeno (%) manufactured by Orthodyne meets the requirements of industrial gas production. It includes the Paramagnetic sensor technology.
Principio
The automatic manufacturing process is capable of very high precision leading to levels of quality and reproducibility beyond anything that is currently available on the market.
With the possibility to set individual measuring ranges, it allows the analyzer to be easily tailored to specific measurement requirements. An outstanding characteristic of this analyzer is its long term stability. In most applications, calibration of the zero-point with ambient air or nitrogen is only required once a month. It is also suitable for measuring rapid changes in the concentration of the sample gas.
Tipo de configuraciones
Rango |
---|
95-100% / 98-100% |
99.5-100% |
0-10% / 0-25% / 70-100% o 85-100% |
Especificaciones Técnicas
Rango de medición | Various configurations possible |
Analog Output signal | 2 x 4..20mA configurable |
Digital Output | Ethernet-10/100BASE-T (standard) 4 x D/O configurable (standard) Modbus RS232/485, Profibus DP/PA and Profinet (optional) |
Linearity | < 0.5% of span |
Repeatability | < 50ppm O2 |
Zero Drift | ≤ 3 % of span of the smallest measurement range per week |
Sensitivity Drift | ≤ 0.1 Vol.-% O2 per week or ≤ 1 % of measured value per week (not cumulative), whichever is smaller. |
Output Fluctuation (2 σ) | ≤ 25 ppm O2 at electronic T90 time (static /dynamic) = 3 / 0 sec |
Detection Limit (4 σ) | ≤ 50 ppm O2 at electronic T90 time (static /dynamic) = 3 / 0 sec |
Warm-up (stability) time | < 1h |
Tiempo de respuesta T90 | T90 ≤ 4 sec at a sample gas flow of 90 l/h and electronic T90 time (static/dynamic) = 3 /0 sec, gas change from N2 to air |
Sample gas T° | +5 to 45°C |
Dew point | 5 °C below the temperature throughout the sample gas path |
Pressure | 2-100 hPa |
Flow rate | 30 - 90 L/h |
Connections | ⅛ NPT female thread (inlet and outlet) |
Alimentación | 100 – 240 VAC (-15% + 10%) 50-60 Hz (+- 3%) |
Dimensiones | Rackable unit 19’’ - Total height: 3U (133mm) – Depth: 365 mm. |