Analizador de trazas de Oxígeno - OZR 8000
Control of pure gases and gas mixtures
General
OZR 8000 Analizador de trazas de Oxígeno meets the requirements of industrial gas production y thermal treatment industries.
Principio
Measuring principle
–The oxygen in the sample gas diffuses from the sample side of the ZrO2 cell through to the outside (ambient), thereby a potential gradient exists which is measured via the inner/outer electrodes.
–The change in the potential gradient is proportional to the change in concentration of oxygen present in the sample gas
–The reference value is always ambient air.
Tipo de configuraciones
Ranges |
---|
Factory preset 0 – 1/10ppm |
Freely configurable up to 250 000 ppm |
Especificaciones Técnicas
Measurement method | Zirconium dioxide detector |
Ranges | Factory preset 0 – 1/10ppm Freely configurable up to 250 000 ppm |
Analog Output signals | 2 x 4..20mA configurable |
Salida Digital | Ethernet-10/100BASE-T (standard) 4 x D/O configurable (standard) Modbus RS232/485 Profibus DP/PA and Profinet (optional) |
Linearity | With such measuring principle, Zirconium Dioxide cells are base linear. |
Repeatability | < 1 % of the measurement range or 100 ppb O2 (whichever is greater) |
Zero Drift | < 1 % of the measurement range per week or 250 ppb O2 (whichever is greater) |
Sensitivity Drift | <1 % of measurement range per week or 250 ppb (whichever is greater) for pure samples. |
Output Fluctuation (2 σ) | < ± 0.5 % of the measured value or 50 ppb O2 (whichever is greater) |
Detection Limit (4 σ) | < ± 1 % of the measured value or 100 ppb O2 (whichever is greater) |
Warm-up time | < 15 minutes |
Response time | T90 < 60 sec at a sample gas flow of 8 l/h and electronic T90 time (static/dynamic) = 3 /0 sec |
Sample gas T° | +5 to 45°C |
Pressure | ≤ 70 hPa, outlet pressure: atmospheric |
Flow rate | 4 – 20 L/h |
Connections | 3mm Swagelok ® (inlets) 1/8 NPT internal thread (outlets) |
Alimentación | 100 - 240 V AC (- 15 %, + 10 %) 50-60 Hz (± 3 Hz). |
Dimensiones | Rackable unit 19’’ - Total height: 3U (133mm) – Depth: 365 mm. |
* For an outlet pressure between 800 y 1750 mbar
Aplicaciones
> Air separation plants
> Specialty gas laboratories
> Process control
> Steel industry
> Cryogenic truck loading station
> Basic research
> Materials research
> Monitoring of atmospheres
Descubra cuál analizador(s) necesario para su solicitud analítica. Consulte la aplicación aquí !
Características
Long sensor lifetime
Trace oxygen measurement in pure gases
Factory preset measuring ranges 0…1/10 ppm
Freely configurable up to 250 000 ppm
Clear-text status messages
Self-monitoring function indicates when maintenance is required